GE Healthcare · CTBrightSpeed26011816818621826011818550103
BrightSpeed HPower, Delight, Velocity or MERI HPower
Symptom
Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits
Error message
Error: 260118168 (40-0105) Rotation Phases unbalanced See LD for error explanation Code 40 Error code: 40-0105H Message/explanation: Rotation Phases unbalanced The amplitude difference of the current between main and auxiliary is too large. Potential cause: 1. Tube stator winding problem 2. Incorrect wiring (one wire open) 3. Phase shift capacitors inverted or wrong value or not connected 4. Rotation board Recommended action/ Troubleshooting guide: Disconnect rotation-to-tube-stator cable and check stator windings impedance. If not OK, replace tube. Check wiring from rotation board to tube. Check that phase shift capacitors are correctly connected Check that phase shift capacitors are not in short-circuit Run Rotation diagnostic (See diagnostic section). If test fails, replace Rotation board Tue May 9 09:26:56 2006 Error: 186218 Host: IRIS Process: tgp_out File: tgp_in_input.c Line: 335 XG Error phase is 3 code is 40. detail error code : 40 - 0105H Rotation error : Rotation Phases unbalanced Error: 260118185 (50-0204) Filament permanent open circuit See Long Description for complete error. Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Je
Applies to
- Models
- BrightSpeed
- Product
- BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 See Long Description for details 2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 Part #: 2246223-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HV TANK HEMIT JEDI 60DC (SPARE) FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required