GE Healthcare · MRDiscovery224857928727224859222485841207812070
SCIM communication error caused by defective SCP3 board
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2248579 A timeout occurred when trying to receive a SCIM packet 2248579 Host: PMR5_OC0 Process: NSP File: SCP:KeyboardLinkResources Line: 471 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Communication Error. A timeout occurred when trying to receive a SCIM packet. Opcode = 0x60 Check cable connection between the SCIM and WIM and cable connection between the SCIM and System Cabinet. 2248592 SCIM Configuration Failed SCIM link errors caused the configuration to fail 2248592 Host: PMR5_OC0 Process: NSP File: SCP:OpWorkspaceAutoConfig Line: 364 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Gradient Echo Coil: HEAD Scan: 28727/1 SCIM Configuration Failed. SCIM link errors caused the configuration to fail. Please fix and perform a TPS Reset 2248584 SCIM Communication Error. Thu Apr 10 07:21:09 2008 2248584 2 NSP SCP:KeyboardLinkHWO 1920 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications SCIM Communication Error. The SCIM failed to respond to a watchdog packet from the SCP. Watchdog packets waiting for a response: 0x0000 Check cable c
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 2294300-22 MGD SCP3 RoHS Compliant board 2294300-22 MGD SCP3 RoHS Compliant board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.