GE Healthcare · CTDiscovery2300000771246877321230000898
Scanner Hardware Stopped Scan.
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Errors:230000077 Unable to open the DIP device driver See long description for error Errors: 230000077 Unable to open the DIP device driver 1246877321 1 1 Mon Jul 6 10:48:41 2009 230000077 6 darc dataacq dataacq_main.cxx 361 No device driver installed Unable to open the DIP device driver. Driver name: /dev/dip or /dev/gemsvdip Open mode: read/write OS error : 2 --- No such file or directory Error: 230000898 Unable to open the DIP device driver. See Long Description for complete error. Error: 230000898 Unable to open the DIP device driver Fri Sep 25 14:22:55 2009 Error: 230000898 File: dataacq_main.cxx Line: 361 Host: darc Process: dataacq OS error : 2 --- No such file or directory Open mode: read/write Driver name: /dev/dip or /dev/gemsvdip Unable to open the DIP device driver. No device driver installed
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5255299-10 HDIG - HD750 Data Aquisition and Image Generator Computer Assembly - used on GOC6 OC See Long Discription for Details Part #: 5255299-10 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HDIG - VCT HD Data Aquisition and Image Generator Computer Assembly FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution