Pular para o conteúdo principal
GE Healthcare · MRSigna2262134

Coil ID Fails and CoilID Light is Red Due to the 2423192 Head Neck Unit 1.5T Brain Spine Coil.

Symptom

Coil ID Fails and CoilID Light is Red See Long Description Customer is attempting to scan, but getting error that something is wrong with coil ID. The Coil ID LED turns red.

Error message

2262134 Coil validation error. Coil ID device was not detected on a coil. Coil ID is required for this connector See Long Description Tue Sep 2 07:14:13 2008 Error: 2262134 Host: hde Process: NSP File: SCP:SRI Line: 348 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 Coil validation error: Coil ID device was not detected on a coil. Coil ID is required for this connector. Please try connecting coil again. Extended Error: This could be a problem with the coil ID communication chain which includes the Coil, the cable between the Coil and the LPCA, the LPCA, the cable between the LPCA and SRI, and the SRI. Please run Coil ID diagnostics.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2423192 HEAD NECK UNIT 1.5T BRAIN SPINE FRU 2423192 HEAD NECK UNIT 1.5T BRAIN SPINE FRU This coil does not ship with the Coil ID PCB. If existing coil has Coil ID PCB then transfer to new coil as per process described in FRU replacement service document. Coils being replaced that do not have Coil ID PCB instead utilize Coil ID circuitry resident inside existing cable and connector assembly which should also be transferred to new coil. See service replacement document for details. 212002950 July 17, 2009. Promoting this service part per 1.5T HNS Coil Program, eNPI # 013865, 212035293, 07 Nov 2008
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues