Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT26010770526010720650180260107204

Gantry does not rotate.

Symptom

Gantry does not rotate. Customer noticed that gantry does not rotate.

Error message

Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. See LD for complete error Error: 260107705 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Mon Mar 14 10:34:57 2005 Error: 260107705 Host: tgp Process: tgp_subsys File: scanbusDriver.c Line: 233 Mon Mar 14 10:34:57 2005 Host : Table Gantry Processor Ermes # : 260107705 Exception Class : Abort Severity : Secondary File : scanbusDriver.c Line# : 233 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Could not send axial drive message in the allotted time. Axial serial link is down. Possible causes: Axial CAN jumper, Axial fuse, interconnects, CAN devices. Retries used: 0

Error: 260107206 Axial Motor Control (AMC) logic will NOT accept commands until the Axial Drive is properly configured/initialized. See Long Description for full error Wed Apr 7 21:04:02 2004 Error: 260107206 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3771 Wed Apr 7 21:03:47 2004 Host : Table

Applies to

Models
LightSpeed, VCT
Product
LightSpeed RT4 WB O G4 G HPT WB O M4 G HPT LightSpeed RT4 with GRE and MDAS4 or GDAS4 gantry At GE Medical Systems, we’re pioneering exciting CT technology for Radiation Therapy planning. Introducing the newest addition to our LightSpeed® series – LightSpeed RT, the first multi-slice, wide-bore CT scanner with the largest field of view in the industry. With a generous 80 cm gantry opening, the LightSpeed RT bore is 10 cm larger than conventional scanners. This provides complete flexibility for patient positioning freedom – even for the most demanding oncology needs. And the 65 cm variable field of view delivers 15 additional centimeters of visual coverage, allowing clinicians to see a patient’s entire anatomy for precise radiation therapy planning. Position and precision Thin, multi-slice imaging is the extra bonus – enabling respiratory gating, providing precise target delineation and delivering higher resolution digitally reconstructed radiographs (DRRs) than ever before. All adding up to the ability for you to deliver what really matters most – better cancer care for your patients. Plus, the LightSpeed RT scanner is available with the Exact™ couch insert, enabling you to consistently replicate patient immobilization for treatment. With LightSpeed RT, the future of your Radiation Oncology department is wide open. Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2218570-2 H-POWER FUSE KIT used on Lightspeed 5x systems. See Long Description for contents of fuse kit. Part #: 2218570-2 Part Type: GP CONSUMABLES Part Weight: 43.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H-POWER FUSE KIT FRU: Yes Status: Orderable Comments: Not Available Fuse KIT contents Mark Number Name Qty Description 1 2351492 2.0 200A FUSE 2 2364059 2.0 GLASS FUSE 3 46-170021P50 2.0 "FUSE 12 AMPS 250 VOLTS BUSSMAN MDA12 DUAL ELEMENT." 4 46-170021P106 2.0 FUSE 8 AMPS 250 VOLTS BUSSMANN ONLY 5 2336517-2 2.0 FUSE 25 AMPS 700 VOLTS 2.5M SECONDS 6 46-170021P52 3.0 "FUSE 3 AMPS 250 VOLTS(REVIEWED TK, 7/91)." 7 46-170021P10 2.0 "006.000A 0250V 3AG FAST UL/C 8 46-170021P101 1.0 "FUSE 20 AMPS 700 VOLTS 2~ LONG X 9/16~ DIA FAST BLOW" 9 46-170021P11 2.0 "7.0A, 250V SLO-BLO FUSE. *TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 10 46-170021P31 2.0 "1/2A, 250V SLO-BLO FUSE. TYPE 3AG, 1.25~ X 0.25~ GLASS BODY. 11 2106993-5 2.0 FUSE 20 AMPS 500 VOLTS .3 SECONDS 12 46-327160P1 2.0 "12.0A, 125VAC, DUAL ELEMENT TIME DLY FNM-12, 0.406 X 1.5 LG, FIBER BODY" 13 2379651 1.0 FUSE - 700 V, 200 A, FAST SEMI
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.