Pular para o conteúdo principal
GE Healthcare · MRExcite22522932252301

Multicoil Bias Short or Open Faults Due to Cable.

Symptom

Unable to Scan With Multi Coil Bias Faults Unable to scan with Multi Coil Bias Faults

Error message

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2252301 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCDx Channel x of y, counting from 1 Value x mA Limit y mA See Long Description for Extended Error 2252301 MRI_GB_OW0 NSP SCP:DCB 816 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 (or MCD2) Channel 1 of 16, counting from 1 Value: 2 mA Limit: 298 mA Mask of failed channels: 0x1 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. An open circuit fault occurs when the actual value is less than the pre-defined limit. Check system cables, connectors, coils and configurations. Re-seat all coil connectors. The mask of all failed channels indicates additional channels with the same fault - convert the hex number to a bitmap to decipher the mask, with channel 1 indicated by the least significant bit and channel 16 indicated by the most significant bit. F

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
5115097 CABLE AND TRACK ASM 5115097 CABLE AND TRACK ASM
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues