Symptom

YMS Tube warmup and cal fails with Warmup SD/Mean is out of range.

Error message

Error: 189057 Error in Vector Check Error in SD/Mean Error in Ratio !! See the LD for the full error code Error: 189057 Error in Vector Check Error in SD/Mean Error in Ratio !! Error: 189057 File: warmup_calc.cxx Error in Vector Check !! Error: 189057 File: warmup_calc.cxx Error in SD/Mean !! Error: 189057File: warmup_calc.cxx Line: 452 Error in Ratio !! Error: 189057 File: warmup_calc.cxx Line: 464 Error in HPF !! !

Error: 189060 Warmup ratio is out of range: 574ch. See Long Description for complete error Thu Jun 2 01:00:01 2005 189060 2 CTO1 OC0 ToolCtrl warmup errchk.c 177 Warmup ratio is out of range: 574ch.

Applies to

Models
HiSpeed
Product
HiSpeed Dual, ProSpeed FII Series HiSpeed Dual, ProSpeed FII Series Product Name: Hispeed Dual ( Linux ) CT/Be2 A list of types. (a Product/hardware table) This console has an HP8000, HP8200, or HP8400 Computer (not GRE console) An On-site / Remote identification aid . (Example: check_config) HP Host computer A Table of options (included, optional, new introductions, …) HiSpeed Dual for China and Global B71402CA has the following as standard Options 1.5 sec fast scan, Power 200ma, Thichness 1mm, Raw data 5000, Image 100000 A list of Software that has run on this product (Product / Software Compatibility Table) With HP8000/8200 : • OS Version 4.5.15144980•Software Version 2.235165193 (except China)•Software Version 2.235172510 (for China only)•Patch 2.245179238•Patch 2.255197828•Patch 2.275267772 With HP8400 : •OS Version 6.2.95265475•Software Version 3.105270457•Patch 3.305267336•DVD Interchange 6.15270187•Patch 3.40 (See Note 8)5316222•DVD Interchange 7.05316345N/ Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links: Drawing Print 2320306-2CCT : http://gemspdm.med.ge.com/ematrix/common/eLibDocument.jsp?type=type_DrawingPrint&name=2320306-2CCT&rev=9 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2138808 bowtie filter 2138808 bowtie filter Part #: 2138808 Part Type: GP CONSUMABLES Part Weight: 3.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9234MW FILTER 2138808 bowtie filter Part #: 2138808 Part Type: GP CONSUMABLES Part Weight: 3.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: P9234MW FILTER Comments: FLS - BOW TIE FILTER FOR YMS HISPEED& ESR CT MS7368 01/10/00
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.