Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260100042260125041130782813083224294967295

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260100042 Cradle was NOT ready See the long description for the full error code Tue May 17 16:02:03 2005 Error: 260100042 Host: tgp Process: tgp_subsys File: scHdwRdy.c Line: 70 Tue May 17 16:02:02 2005 Host : Table Gantry Processor Ermes # : 260100042 Exception Class : Soft Severity : Sec/Soft File : scHdwRdy.c Line# : 70 Function : No System Function Reported Scan Type : Scout Scan: 106/1/1 Cradle was NOT ready

Error: 260125041 Motion without command in Cradle drive See Long Description for an example error message. Tue Aug 21 12:19:15 2007 Error: 260125041 Host: tgp Process: tgp subsys File: ta callback.c Line: 507 Tue Aug 21 12:19:14 2007 Host : Table Gantry Processor Ermes # : 260125041 Exception Class : Abort Severity : Primary File : ta callback.c Line# : 507 Function : No System Function Reported Scan Type : None/Unknown Scan: 4747/3/0 Motion without command in Cradle drive Current pos = 1307828 Last stop pos = 1308322 Tolerance = 250 Time since last stop = 20 Last target pos = 4294967295 Hardware status = 8104

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5122207-2 ABSOLUTE ENCODER FOR CRADLE GT (replaces 5122207) See long description for details Part #: 5122207-2 ABSOLUTE ENCODER FOR CRADLE GT (replaces 5122207) Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: ABSOLUTE ENCODER FOR CRADLE GT FRU: Yes Status: Orderable Comments: This part is used in GT-1700mm and GT-2000mm tables. Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.