Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23001708850019230023033

LightSpeed 4.X - GOC2

Symptom

Fastcal fails the mylar window check. Fastcal fails the mylar window check

Error message

Error: 230017088 The mylar window check failed See Long Description for complete error Mon Aug 23 05:50:31 2004 Error: 230017088 Host: usafaultra Process: accs File: aircal processing.c Line: 1025 The mylar window check failed - Scan 50019/1/3

Error: 230023033 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts. See Long Description for complete error Sat Jan 8 08:06:05 2005 230023033 Host: uhct Process: dailyPrepRx File: DailyPrepUtils.c Line: 2967 Mylar Window Check has failed. Mylar Window may be dirty. Calibration vectors generated with the dirty mylar window may cause Image artifacts.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5142849-4 SCAN WINDOW LIGHTSPEED SERIES H2 THRU HP RIM COVERS (replaces 2101967-4). See long description for details. Part #: 5142849-4 Part Type: GP CONSUMABLES Part Weight: 7.94 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASSEMBLY, SCAN WINDOW, LIGHTSPEED SERIES H2 THRU HP RIM COVERS, WEATHERSTRIP UPDATED MAY 2005 FRU: Yes Status: Orderable Comments: This part is replacement for 2101967-4, please discontinue use of 2101967-4 and use only 5142849-4. Added per ECR2014906, ms13452 050713
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues