Pular para o conteúdo principal
GE Healthcare · MROptimaBrivo2270016

Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2270016 ADS1 VRF did not receive correct device ID from RRX module in scan room. RRX does not exist or is not connected to VRF. The VRF is located at ICN1(ICN2), Fri Feb 27 07:06:54 2009 Error: 2270016 Host: mr1 heri Process: LxMxConsumer File: VrfTrace.cpp Line: 80 Server Name: ADS1 VRF did not receive correct device ID from RRX module in scan room. RRX does not exist or is not connected to VRF. The VRF is located at ICN1(ICN2), if ADS(ADS1) reports this error. VRF ID: 0 Device ID: 2-

Applies to

Models
Optima, Brivo
Product
Optima MR360 Alias; HDsv. This concept covers both Optima MR360 and Brivo MR355. Both products are produced in Beijing, China.
Software
SV20.1 Applies to Optima MR360 and Brivo MR355, a.k.a. HDsv. Both products are made in Beijing, China.
Component
5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.