Pular para o conteúdo principal
GE Healthcare · MROptima

ICN Fails to Configure Due to Sun ICN VRE.

Symptom

ICN VRE configuration fails

Error message

ERROR DHCP address requests failed.Only received 0 of 1 DHCP requests after OS install. Error found in /usr/local/bin/install/log/VRE_Install.log during Sun ICN VRE configuration.ERROR DHCP address requests failed.Only received 0 of 1 DHCP requests after OS install. Please check ethernet cables, reseat connections and retry VRE configuration

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5337894 Image Compute Node - DV 16CH GEN 3 5337894 Image Compute Node - DV 16CH GEN 3
Area
MR