Pular para o conteúdo principal
GE Healthcare · MRExcite2262134

Head Coil will Not Stay Connected Due to Damaged Pins in the Head T/R Assembly.

Symptom

Cannot scan with coil validation error. Cannot scan with error coil validation error

Error message

2262134 Coil validation error. Coil ID device was not detected on a coil. Coil ID is required for this connector See Long Description Tue Sep 2 07:14:13 2008 Error: 2262134 Host: hde Process: NSP File: SCP:SRI Line: 348 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 Coil validation error: Coil ID device was not detected on a coil. Coil ID is required for this connector. Please try connecting coil again. Extended Error: This could be a problem with the coil ID communication chain which includes the Coil, the cable between the Coil and the LPCA, the LPCA, the cable between the LPCA and SRI, and the SRI. Please run Coil ID diagnostics.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2296092-3 Head TR Assembly 1.5T Excite-III or Excite HD Head TR Assembly 1.5T Excite-III 2296092-3 Head TR Assembly 1.5T Excite-III 2296092-3
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues