GE Healthcare · MRDiscovery2267835

gate-drive over-current Errors Due to XGA Amplifier.

Symptom

Will Not Scan with Gradient Errors.

Error message

2267835 A gate-drive over-current has occurred on % axis XGA Aug 18 Tue 05:04:17 2267835 NSP SCP:FspMac.cpp 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 A gate-drive over-current has occurred on Z axis XGA. Extended error log: 2267835ERROR: EM_XGD_AS_GD_OC A gate-drive over-current has occurred on %b XGA. If this message persists after TPS resets, the corresponding XGA should be replaced.

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER