Symptom

Scanner failed to scan with Image Allocation errors See Long Description for error examples Scanner failed to scan with Error: 2021 mkdir failed /usr/g/sdc_image_pool Input/output error and Error: 210000434 IMAGE ALLOCATION has returned null linkedTuple or null linkedTuple->data.

Error message

Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToScanRequest function. Nov 22 08:55:38 2006 Error: 200107023 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 148 Function: Data Acquisition : OC Processing srx_sm_event_notify: A Failure had been detected while calling srx_sm_generate_action function. Tue Aug 14 16:28:25 2007 Error: 200107023 Host: cccgvct1 Process: scanRx File: srx_sma_manage_scan.c Line: 733 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling SIRx_areScanFilesAvailable function.

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Software
09MW08.10 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) June 2009 Inital M3 release. This software is now obsolete, this site should be upgraded to 11HW12.5 software released via FMI 25427 09MW08.10 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) June 2009 Inital M3 release. LightSpeed 7.2 ( 09MW08.10 ) Software Release supports GOC6 consoles used on VCT 64, VCT XT, VCT XTe and VCT Std. - LightSpeed 7.2 (09MW08.10 )Software for GOC6 Collector 5328048 Rev 3 Includes OS and Application CD's - Operating System Linux 5.3.8 Software CD 5323584 Rev 1 - Application LightSpeed VCT GOC6 DVD 5336533 Rev 3 - Class C LightSpeed 7.2 5336535 Rev 2 - Class M LightSpeed 7.2 5336536 Rev 2 An On-site/Remote identification aid . (Example: check_config) Result from check_config or swhwinfo 09MW08.10.V40_H_H64_G_GTM 09MW08.10.V40_H_H64_G_GTM Please create an ITRAK report for any software related issues
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.