Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300110755000082000102000230011008468000

Scanning aborted with scan data errors

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230011075 Unable to open a scan file on the scan data disks. Read of scan file header failed See the LD for error codes Error: 230011075 Unable to open a scan file on the scan data disks. Read of scan file header failed in read Thu Dec 29 07:56:06 2005 Error: 230011075 Host: srct Process: scanmgr File: sdio_xfs_io_ops.c Line: 662 Function: Data Acquisition : SBC Processing Scan Type: Axial Scan: 50000/8/13 Scan Seq Id: 82000 Unable to open a scan file on the scan data disks. Read of scan file header failed in read() call. Scan: CT01/50000/8/13 Bytes to read: 198h Starting offset: 0h Number read: 0h OS error : 0 --- Success Or Fri Oct 6 08:27:13 2006 Error: 230011075 Host: ct01 Process: scanmgr File: sdio_xfs_io_ops.c Line: 662 Function: Data Acquisition : SBC Processing Scan Type: Helical Scan: 5144/3/1 Scan Seq Id: 102000 Unable to open a scan file on the scan data disks. Read of scan file header failed in read() call. Scan: CT01/5144/3/1 Bytes to read: 198h Starting offset: 0h Number read: ffffffffh OS error : 11 --- Resource temporarily unavailable

Error: 230011008 Unable to remove scan file from scan data disks. Error: 230011008 Unable to remove scan file from scan data

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2362873 SDDA (Scan data disk array) used on GOC3. See Long Description for details. The GOC3 to GOC4 upgrade kits are now under 1 part number, 5144058-500 plus the FE will have to also order the DARC2 that is needed by the DAS as a separate order. Part #: 2362873 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SCAN DATA DISK ARRAY - GRE FRU: Yes Status: Orderable Comments: OBSOLETE PART case 8391074 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 05-23-07 ** FE ASSIST ** For CT not PET.. IF no stock of 2362873 SDDA must upgrade to DARC II + ICOM Kit per one of the following part number's * 5144058-500, 16 MDAS 16,8,4 GDAS DARC Replacement kit, JARRELL * 5144058-501, 4,8 MDAS DARC Replacement Kit, JARRELL * 5144058-502, 4 SDAS DARC Replacement Kit, JARRELL Documentaion 5194645-100 Must order software release 06MW29.7, p/n 5193989 and perform a LFC to support this FRU if system is not currently running 06MW29.7 or better (2/19/07 212001261) *** 2ND OPINION NOTES *** Look at the DARC messages log for SCSI disk errors located at /var/log/messages
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required