Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1855485052511833618331634867

LightSpeed 4.X - GRE

Symptom

CTx00088 Proactive svc call by PDS See long description for details CTx00088 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service

Error message

Error: 185548 The HEMRC operating frequency is below the minimum value. See Long Description for full error Error: 185548 Suite: CT01 Host: OBC Process: Diag Exec File: hemrc_manager.c 1.5 Method: gen_rotor Line: 859 Function: Diagnostics XRAY GENERATION : Tube Rotor Control Test Name: HEMRC Manual Unique ID: 50525 Iteration: 1 Exc. Level: Hard Time: 118336 Error Type: Hardware The HEMRC operating frequency is below the minimum value. Actual frequency: 219.172 Minimum frequency limit: 249.995

Error: 183316 (The HEMRC failed to reach the Run speed in the allotted time) Open Long Description for more details Tue Aug 10 17:58:58 2004 Error: 183316 Host: OBC Process: Generator File: hemrc_manager.c 1.14 Function: XRAY GENERATION : Tube Rotor Control Scan: 34867/1/1 Type: Scout Exception Level: Pri/Most Severe The HEMRC failed to reach the Run speed in the allotted time

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2115199 HEMRC - HIGH EFF MTR ROTOR CTRL used on LightSpeed 2.x (Plus), 3.x (Ultra) and 4.x (16 Slice) Systems See Long Description for more info and picture Part #: 2115199 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 7.28 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HIGH EFF MTR ROTOR CTRL FRU: Yes Status: Orderable Comments: FLS - HIGH EFF MTR ROTOR CTRL PURCH SPEC USED ON FAST RHAPSODE 7/1/97 M3 FW28 GWENN 08992 THIS IS THE AC DRIVE FOR HEMRC THIS IS OUT-SOURCED TO A 3RD PARTY FOR ALLEN-BRADLEY. NOT DO NOT DISCOUNT - PER SHAKUN 10/30/97 GWENN MS08992 6/02 BOT FROM HOLT ELECTRIC A/B P#: 1336SBRF50-AN-EN901-MX1-L4 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required