GE Healthcare · MRExcite22522391121430417222381722612392218661
TPS Reset Fails with error 2261239 due to ASC Power Supply.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2252239 Startup Timeout was detected on the Universal Power Monitor 2. 1121430417 3 1 Fri Jul 15 12:26:57 2005 2252239 2 lx-mr NSP SCP:CanAutoConfig 2576 Startup Timeout was detected on the Universal Power Monitor 2.-> EN 168 SR 259 2252239 Startup Timeout was detected on the Universal Power Monitor 1. Fri Jul 15 12:26:57 2005 2252239 2 lx-mr NSP SCP:CanAutoConfig 2576 Startup Timeout was detected on the Universal Power Monitor 1.-> 2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 2261239 ASC NB IF Board was not detected but is configured. See Long Description Fri Jul 15 11:46:26 2005 2261239 lx-mr NSP SCP:CanAutoConfig 2587 ASC NB IF Board was not detected but is configured in MRconfig.cfg. If this board is present, verify power, RF Amp and CAN communication and perform a TPS Reset. If the SSM is present, set usssType=no in MRconfig.cfg and perform a TPS Reset.-> 2252239 Startup Timeout was detected on the RF NB Amp Interface. Fri Jul 15 11:46:26 2005 2252239 2 lx-mr NSP SCP:CanAutoConfig 2576 Startup Timeout was detected on the RF NB Amp Inter
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2363040 COMPACT PCI 3U 200 WATT POWER SUPPLY 2363040 COMPACT PCI 3U 200 WATT POWER SUPPLY
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.