Pular para o conteúdo principal
GE Healthcare · CTLightSpeed200002714200000436200002712200002744

LightSpeed 4.X - Octane

Symptom

Scout images fail to reconstruct. See long description for details Scout images fail to reconstruct. Rebooting, cycling console power, etc. does not help.

Error message

Error: 200002714 MFM Error : Format failure DICOM image not created See Long description forfull error Fri May 27 09:40:15 2005 Error: 200002714 Host: ct2 Process: imagecreate File: ic_imagecreate_req.c Line: 492 Image : 4243/1/1 Recon Sequence Id : 50 MFM Error : Format failure DICOM image not created

Error: 200000436 Image not installed because image available event flagged image as not being available. Fri May 27 09:40:15 2005 Error: 200000436 Host: ct2 Process: dms File: InstallImage.cxx Line: 1113 Image 4243/1/1 not installed because image available event flagged image as not being available.

Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies] See long description for full error. Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies]. The variable information gives the failing Group and Element and shows the failing string. Fri May 27 09:40:15 2005 Error: 200002712 Host: ct2 Process: imagecreate File: ic_mfm_main.c Line: 448 Image : 4243/1/1 Recon Sequence Id : 50 MFM error : Format failure Group Number : 40 Element 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
405I.3_H4.0M5.1 for LightSpeed 4.x LS2002M5.1 405I.3_H4.0M5.1 for LightSpeed 4.X LS2002M5.1 (IRIX) LS2002M5.1 Supports New Motorola-bd P/N 2370825. FRU Collector 5114901 (Motorola-bd 2370825 + Instructions 5114494-100) will be available around FW30, 2004 GOC1 Console OPERATING SW: Software Collector 5112672 Patch C.1 5113667 CLASS C: 2395172 or 2401759 (6 pack) 1104I.2_H4.0M5CLASSC CLASS M: 2395173 or 2401759 (6 pack) 1204I2_H4.0M5CLASSM 405I.3_H4.0M5.1 for LightSpeed 4.X LS2002M5.1 (IRIX) LS2002M5.1 Supports New Motorola-bd P/N 2370825. FRU Collector 5114901 (Motorola-bd 2370825 + Instructions 5114494-100) will be available around FW30, 2004 GOC1 Console OPERATING SW: Software Collector 5112672 Patch C.1 5113667 CLASS C: 2395172 or 2401759 (6 pack) 1104I.2_H4.0M5CLASSC CLASS M: 2395173 or 2401759 (6 pack) 1204I2_H4.0M5CLASSM 5112672 LIGHTSPEED 16 MDAS M5.1 Collector Number Quantity Description/Name, 5112617 1.0 PC LS2002M5.1 FOR H4 5112618 1.0 PC LS2002M5.1 FOR H4 APPS DISK 5112619 1.0 PC LIGHTSPEED PATCH DISK
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.