GE Healthcare · MRSigna22445732245044924192244868

Defective HFA - High Fidelity Amplifier.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244573 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage Error: 2244573 Host: WMR2 Process: NSP File: SCP:Grad Line: 2587 The GP reported a Gradient Amplifier X Axis Internal Wiring Fault or Control Undervoltage.->

2245044 The GP reported a Gradient Amplifier Y Axis Internal Wiring Fault or Control Undervoltage. Fidelity Test see long description When Fidelity diagnostics are run the following error is reported. 2245044 Diagnostic: Fidelity Test Iteration: 1 of 1 Action: 1Task ID: 0xffc3190 Time: 92419 Object: SCP Test Task Function: Gradient: N/A GRADIENT DRIVER TESTS - The GP reported a Gradient Amplifier Y Axis Internal Wiring Fault or Control Undervoltage. Fidelity Test-> Extended error message-> Swap Gradient Amplifier Control Boards with another axis to isolate Gradient Amplifier vs. Gradient Amplifier Control Board.

2244868 The GP reported multiple Gradient Amplifier %Axis Internal Wiring Faults or Control Undervoltages 2244868 Host: lx-mr Process: NSP File: SCP:Grad Line: 2613 The GP reported multiple Gradient Amplifier % Axis Internal Wiring Faults or Control Undervoltages.-

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion: HFA with FRU Package and Lift Brackets 2393906-4
Area
MR

Full cause & step-by-step resolution