GE Healthcare · MRSigna22445732245044924192244868
Defective HFA - High Fidelity Amplifier.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244573 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage Error: 2244573 Host: WMR2 Process: NSP File: SCP:Grad Line: 2587 The GP reported a Gradient Amplifier X Axis Internal Wiring Fault or Control Undervoltage.-> 2245044 The GP reported a Gradient Amplifier Y Axis Internal Wiring Fault or Control Undervoltage. Fidelity Test see long description When Fidelity diagnostics are run the following error is reported. 2245044 Diagnostic: Fidelity Test Iteration: 1 of 1 Action: 1Task ID: 0xffc3190 Time: 92419 Object: SCP Test Task Function: Gradient: N/A GRADIENT DRIVER TESTS - The GP reported a Gradient Amplifier Y Axis Internal Wiring Fault or Control Undervoltage. Fidelity Test-> Extended error message-> Swap Gradient Amplifier Control Boards with another axis to isolate Gradient Amplifier vs. Gradient Amplifier Control Board. 2244868 The GP reported multiple Gradient Amplifier %Axis Internal Wiring Faults or Control Undervoltages 2244868 Host: lx-mr Process: NSP File: SCP:Grad Line: 2613 The GP reported multiple Gradient Amplifier % Axis Internal Wiring Faults or Control Undervoltages.-
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion: HFA with FRU Package and Lift Brackets 2393906-4
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.