Guided Install Hardware Tab had the Spectroscopy RF Amplifier selected with No Spectroscopy RF Amplifier Installed in th…

Guided Install Hardware Tab had the Spectroscopy RF Amplifier selected with No Spectroscopy RF Amplifier Installed in the system.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2261240 ASC BB IF Board was not detected but is configured in MRconfig.cfg. 1121427986 3 1 Fri Jul 15 11:46:26 2005 2261240 2 lx-mr NSP SCP:CanAutoConfig 2597 ASC BB IF Board was not detected but is configured in MRconfig.cfg. If this board is present, verify power, RF Amp and CAN Communication and perform a TPS Reset. If the SSM is present, set usssType=no in MRconfig.cfg and perform a TPS Reset.-> EN 351 SR 162

2252239 Startup Timeout was detected on the RF BB Amp Interface. 1121427986 2 1 Fri Jul 15 11:46:26 2005 2252239 2 lx-mr NSP SCP:CanAutoConfig 2576 Startup Timeout was detected on the RF BB Amp Interface.-> EN 162 SR 351

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR