Fails TPS Reset when booting up to Applications or when Resetting TPS due to XGD Power Supply.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2256925 StartupCondition error on XGD StartupCondition resulted in TPS reset failure See Long Description 2256925 SLMAINMR LxMxConsumer services/initialization/ServerInitManager 767 Server Name: SC StartupCondition error on XGD StartupCondition resulted in TPS reset failure.

2267810 Safe mode startup of the XGD (XPS-SB Y axis) failed. Check EMO reset and perform a TPS reset. See Long Description 2267810 "Safe mode startup of the XGD (XPS-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative.->"

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.