GE Healthcare · MRDiscovery22569252267810
Fails TPS Reset when booting up to Applications or when Resetting TPS due to XGD Power Supply.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2256925 StartupCondition error on XGD StartupCondition resulted in TPS reset failure See Long Description 2256925 SLMAINMR LxMxConsumer services/initialization/ServerInitManager 767 Server Name: SC StartupCondition error on XGD StartupCondition resulted in TPS reset failure. 2267810 Safe mode startup of the XGD (XPS-SB Y axis) failed. Check EMO reset and perform a TPS reset. See Long Description 2267810 "Safe mode startup of the XGD (XPS-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative.->"
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.