Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230000030362502560005490003604506230011008

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors

Error message

Error: 230000030 Unable to write offset view data to the scan file. See LD for complete error Error: 230000030 Unable to write offset view data to the scan file Thu Sep 7 12:25:05 2006 230000030 4 ice dataacq dataacq_save.cxx 669 Function: Data Acquisition : Scan Data Save Scan Type: Axial Scan: 36250/200/2 Scan Seq Id: 256000 Unable to write offset view data to the scan file. Bytes being saved: 200000h Save attempt: 649 Last view buffer? 0 Scan aborted. Wed May 4 16:26:10 2005 Error: 230000030 Host: ice Process: dataacq File: dataacq_save.cxx Line: 720 Function: Data Acquisition : Scan Data Save Scan Type: Helical Scan: 2800/3/1 Scan Seq Id: 549000 Unable to write offset view data to the scan file. OS error : 140 --- Remote I/O error Scan file may be deleted manually.

errnoGet () == 3604506 found in the iceConsole.log errnoGet () == 3604506 This is found in the iceConsole.log

Error: 230011008 Unable to remove scan file from scan data disks. Error: 230011008 Unable to remove scan file from scan data disks Error: 230011008 Unable to remove scan file from scan data disks. Wed May 4 16:25:48 2005 Error: 230011008 Host: gsgect2 Process: scanmgr File: sdio_xfs_io_ops.c Line: 1164 Func

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
Area
CT

Full cause & step-by-step resolution