Symptom

Console fails to boot up. The boot hangs with the HP Invent screen displayed on each monitor.

Error message

SCSI error : return code = 0x8000002, Deferred sdb: sense key Hardware Error, Buffer I/O error on device sdb11 See LD for more of the Errors OC: /var/log/messages Jul 6 09:02:10 gectsp kernel: SCSI Error: (0:0:0) Status=02h (CHECK CONDITION) Jul 6 09:02:10 gectsp kernel: Key=4h (HARDWARE ERROR); FRU=02h Jul 6 09:02:10 gectsp kernel: ASC/ASCQ=19h/00h "" Jul 6 09:02:10 gectsp kernel: CDB: 2A 00 08 44 EF 1F 00 00 30 00 Jul 6 09:02:10 gectsp kernel: SCSI error : <2 0 0 0> return code = 0x8000002 Jul 6 09:02:10 gectsp kernel: Deferred sdb: sense key Hardware Error Jul 6 09:02:10 gectsp kernel: Additional sense: Defect list error Jul 6 09:02:10 gectsp kernel: end_request: I/O error, dev sdb, sector 138735391 Jul 6 09:02:10 gectsp kernel: Buffer I/O error on device sdb11, logical block 8843531 Jul 6 09:02:10 gectsp kernel: lost page write due to I/O error on sdb11 Jul 6 09:02:39 gectsp kernel: SCSI Error: (0:0:0) Status=02h (CHECK CONDITION) Jul 6 09:02:39 gectsp kernel: Key=3h (MEDIUM ERROR); FRU=00h Jul 6 09:02:46 gectsp kernel: ASC/ASCQ=0Ch/02h "" Jul 6 09:02:46 gectsp kernel: CDB: 2A 00 08 44 EF 27 00 00 28 00 Jul 6 09:02:46 gectsp kernel: *********************************************

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5117866-5 73 GB HDD used in HPxw8200 GRE. See Long Discription for Details. Part #: 5117866-5 Part Type: GP CONSUMABLES Part Weight: -139998.60 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB HDD for Lighting System Controller 5117866-2 FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.