GE Healthcare · MRSigna2244589
SGA Power Supply Internal Wiring Fault or Control Undervoltage Due to SGA Power Supply.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244589 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage. 2244589 Host: tradmri01 Process: NSP File: SCP:Grad Line: 2903 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage.-
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 2317092-2 SGA ACGD PS ASM W/FRU PACKAGING FOR ACGD ONLY This SGA ACGD Power Supply can only be used on Excite (11x) or earlier systems that use the ACGD gradient subsystem. It cannot be used on Excite HD (12x) or Excite HDx (14x) systems that use the HFD or HFD-S gradient subsystem. 2317092-2 SGA ACGD PS ASM W/FRU PACKAGING FOR ACGD ONLY This SGA ACGD Power Supply can only be used on Excite (11x) or earlier systems that use the ACGD gradient subsystem. It cannot be used on Excite HD (12x) or Excite HDx (14x) systems that use the HFD or HFD-S gradient subsystem.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.