TPS reset fails due to failed PHPS2.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2268913 The MCPDB in the PHPS has detected a power failure. May 16 Mon 09:49:55 2268913 NSP SCP:SriManager 3925 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 The MCPDB in the PHPS has detected a power failure->

2268871 The dock power cable is not detected at the PHPS. May 16 Mon 09:49:55 2268871 NSP SCP:SriManager 3889 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 The dock power cable is not detected at the PHPS.->

2274510 CFB detected that the PHPS is too hot. May 16 Mon 09:49:55 2274510 NSP SCP:CfbHardware 665 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 CFB detected that the PHPS is too hot. Bad cable connection to PHPS or bad PHPS is the likely cause.

2274516 Serial Communication between CFB and MCPDB inside the PHPS is down. May 16 Mon 09:49:55 2274516 NSP SCP:SystemCanInterf 1215 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 Serial Communication between CFB and MCPDB in

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
5167111 Assembly Patient Handling Power Supply 2 for DVMR 5167111 Assembly, Patient Handling Power Supply 2
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.