Scanning aborts with collimator filter homing problem.

Symptom

Scanning aborts with collimator filter homing problem. Scanning aborts with collimator filter homing problem

Error message

Error: 184162 Axial: Ok Collimator: Timeout Table: Ok Generator: Ok SARQ: Ok See LD for full error Error: 184162 Axial: Ok Collimator: Timeout Table: Ok Generator: Ok SARQ: Ok Sat Dec 16 04:33:57 1995 Error: 184162 Suite: GECT Host: STC Proc: Scan Control File: scReRefMgr.c 1.22 Method: No Method Line: 222 951216 0446 03831 Function: No System Function Reported Scan: 3102/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 04:33:56:985 Log Series: 172 Error detected in end-of-series operations. HW/SW Error: Possible hardware (reference) or software error. Status 0x4d: Axial: Ok Collimator: Timeout Table: Ok Generator: Ok SARQ: Ok

Error: 183411 Invalid state of the Filter Switches. See Long Discription for Details Thu May 8 08:15:21 2008 Error: 183411 Suite: ONC1 Host: OBC Process: Collimator File: col_filter.c 1.16 3/3/99 Method: No Method Line: 221 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 2835/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 09:15:21:255 Log Series: 179 Invalid state of the Filter Switches. Hardware status = 0x40 (Filter Switches are BIT1 and BIT2)

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution