Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed260125041900824897741749120

BrightSpeed Select, Delight Select or MERI Select

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260125041 Motion without command in Elevation drive See LD for more details Wed Jun 14 12:48:32 2006 Error: 260125041 Host: tgp Process: tgp_subsys File: ta_callback_gt.c Line: 508 Wed Jun 14 12:48:31 2006 Host : Table Gantry Processor Ermes # : 260125041 Exception Class : Abort Severity : Primary File : ta_callback_gt.c Line# : 508 Function : No System Function Reported Scan Type : Helical Scan: 412/4/1 Motion without command in Elevation drive Current pos = 900824 Last stop pos = 897741 Tolerance = 3000 Time since last stop = 749120 Last target pos = 1 Hardware status = 0

Applies to

Models
BrightSpeed
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5175348 Table Elevation Pot for NP Lite - used on Some Brightspeed and Optima Part #: 5175348 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: Elevation POT For NP lite FRU: Yes Status: Orderable Comments: added per to BrightSpeed Lite RSPL by sso#305010201
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required