Pular para o conteúdo principal
GE Healthcare · CTLightSpeed200000424200000416200000415

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Unable to access images in Image Works System will not scan and unable to access images in Image Works

Error message

Error: 200000424 Image available space failed. See the long description for the full error Sat Jan 22 18:25:36 2005 Error: 200000424 Host: chr09bc1cti Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. code: 200000416 Or Sat Oct 28 06:55:39 2006 Error: 200000424 Host: ct02 Process: dms File: AvailSpace.cxx Line: 155 Image available space failed. Error code: 200000415

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-78UU or 2169940-78 Octane/Octane II Computer 18 GB Disk Drive. See long description Part #: 2169940-78UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Not Available Status: Orderable Stan Pearson, Asset Mgmt move to used material program 6-27-05 Part #: 2169940-78 Part Type: GP CONSUMABLES Part Weight: 1.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: OCT2 18GB 10K RPM DRIVE FRU: Yes Status: Orderable Comments ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 0711261b. (212001261 8/20/08)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.