Pular para o conteúdo principal
GE Healthcare · Nuclear MedicineInfinia

Gantry lateral motion jerky generating terrible noise

Symptom

Gantry lateral motion jerky generating terrible noise The lateral belt is loose and disturbs the close loop speed regulation. The SEM board reacts by sending jerky power to the motor, generating terrible noise. This has nothing to do with mechanical damage or wear... Just belt tension issue... The lateral belt is loose and disturbs the close loop speed regulation

Applies to

Models
Infinia
Product
Infinia I Infinia I Acquisition System Infinia II Infinia II Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
Component
2211588 Lateral timing belt, Infinia/VG 2211588 Lateral timing belt, Infinia/VG 2211588 Lateral timing belt, Infinia/VG
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required