TR Driver Faults Detected due to Body TR cable between J7 on the Driver Module and J10 on the RF amp.

Symptom

Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.

Error message

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2252305 TR Driver Fault Detected Open Circuit Source of Fault is Body TR. Thu Aug 5 11:00:38 2004 Error: 2252305 Host: fmri3te Process: NSP File: SCP:DCB Line: 1197 TR Driver Fault Detected - Open Circuit. Source of Fault is Body TR. Value: 16 mA Limit: 2400 mA->

2224666 Auto Prescan failed, Driver Module Error Occurred. Wed Apr 21 20:00:24 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1334 Auto Prescan failed, Driver Module error occurred.

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution