GE Healthcare · CTLightSpeedVCT2300000771246877321
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cannot perform any scanning -Scan resource acquisition failed. Cannot perform any scanning -Scan resource acquisition failed
Error message
Errors:230000077 Unable to open the DIP device driver See long description for error Errors: 230000077 Unable to open the DIP device driver 1246877321 1 1 Mon Jul 6 10:48:41 2009 230000077 6 darc dataacq dataacq_main.cxx 361 No device driver installed Unable to open the DIP device driver. Driver name: /dev/dip or /dev/gemsvdip Open mode: read/write OS error : 2 --- No such file or directory
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5255299-20 VDIG VCT Data Aquisition and Image Generator Computer used on GOC6 5255299-20 VDIG VCT Data Aquisition and Image Generator Computer used on GOC6 Part #: 5255299-20 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDIG - VCT Data Aquisition and Image Generator Computer Assembly FRU: Yes Status: Orderable Part #: 5255299-20 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VDIG - VCT Data Aquisition and Image Generator Computer Assembly FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.