GE Healthcare · CTLightSpeed200002744200000436200002712200002714
LightSpeed 4.X - Octane
Symptom
Scout images fail to reconstruct. See long description for details Scout images fail to reconstruct. Rebooting, cycling console power, etc. does not help.
Error message
Error: 200002744 Imagecreate traceback Fri May 27 09:40:15 2005 Error: 200002744 Host: ct2 Process: imagecreate File: ic_mfm_main.c Line: 255 Imagecreate traceback Error: 200000436 Image not installed because image available event flagged image as not being available. Fri May 27 09:40:15 2005 Error: 200000436 Host: ct2 Process: dms File: InstallImage.cxx Line: 1113 Image 4243/1/1 not installed because image available event flagged image as not being available. Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies] See long description for full error. Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies]. The variable information gives the failing Group and Element and shows the failing string. Fri May 27 09:40:15 2005 Error: 200002712 Host: ct2 Process: imagecreate File: ic_mfm_main.c Line: 448 Image : 4243/1/1 Recon Sequence Id : 50 MFM error : Format failure Group Number : 40 Element : 254 Error status : -114 Error String : [ bad syntax for SH or LO or AT ] Error: 200002714 MFM Error : Format failure DICOM image not crea
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405I.2_H4.0M5 for LightSpeed 4.X (LightSpeed 16) with H16 Console (GOC1)(2341923, 2341923-2) - showprods returns 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395173 1204I.2_ H4.0M5 CLASSM LightSpeed/ HiSpeed QX/I Class M Software 2395176 2395170, 2395171, 2399867 LightSpeed 16 Applications Software (Disk 1, Disk 2 and Patch A) 405I.2_H4.0M5 FW47-03 LS2002M5. IRIX only. Class C/M Common w/ LS 2X, 3X, & 4X.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required