Symptom

Scout images fail to reconstruct. See long description for details Scout images fail to reconstruct. Rebooting, cycling console power, etc. does not help.

Error message

Error: 200002714 MFM Error : Format failure DICOM image not created See Long description forfull error Fri May 27 09:40:15 2005 Error: 200002714 Host: ct2 Process: imagecreate File: ic_imagecreate_req.c Line: 492 Image : 4243/1/1 Recon Sequence Id : 50 MFM Error : Format failure DICOM image not created

Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies] See long description for full error. Error: 200002712 MFM error: Format failure Group Number : (varies) Element : (varies) Error status : (varies)Error String : [ varies]. The variable information gives the failing Group and Element and shows the failing string. Fri May 27 09:40:15 2005 Error: 200002712 Host: ct2 Process: imagecreate File: ic_mfm_main.c Line: 448 Image : 4243/1/1 Recon Sequence Id : 50 MFM error : Format failure Group Number : 40 Element : 254 Error status : -114 Error String : [ bad syntax for SH or LO or AT ]

Error: 200000436 Image not installed because image available event flagged image as not being available. Fri May 27 09:40:15 2005 Error: 200000436 Host: ct2 Process: dms File: InstallImage.cxx Line: 1113 Image 4243/1/1 not ins

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console 405L.8_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) ConsoleClass C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2394897) 1104L.3_ H4.2CLASSC LightSpeed/ HiSpeed QXi Class C Software M4 5111370 5111371 LightSpeed 4.X Applications Software 405L.8_H4.2M4 FW11-04 FMI 25338
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues