GE Healthcare · MRExcite2220007

CoilID Light is Off or Dark Due to coil plug had screw partially backed out.

Symptom

CoilID Light is Off or Dark with 2220007 The prescribed RF coil and detected RF coil do not match 2220007 The prescribed RF coil and detected RF coil do not match. Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:0, Lower coil code: Null, Rx coil name:HEAD, Rx coil code:bv523rifp0t9r4h3pgqceh64jh7d2n2d

Error message

2220007 The prescribed RF coil and detected RF coil do not match. See long description on reading/interpreting this type of error message Note: An example of this error message, with explanation, is provided below. Wed Mar 3 08:18:58 2004 Error: 2220007 Host: lx-mr3 Process: scn File: RxMgrDownload.m Line: 2074 The prescribed RF coil and detected RF coil do not match. 040303104912225 Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:1, Lower coil code:q8dCUW0PD2EaETIplAp90PLrACM6CW38, Rx coil name:MRI 8HRKNRR PA, Rx coil code:q8dCUWOPD2EaETIpIAp9oPLrACM6CW38 NOTE: Upper coil connector status refers to connector plugged into upper port at "dog house" (i.e. head coil or Med' Adv' QUAD KNEE ("chimney") coil). Lower coil connect status refers to all PA coils / any coil using the lower port at the "dog house". Coil code should match that of what was detected (in this case the zeros were config'd as O and o (oh's - upper and lower case), and the number one following "ETIp" was an "I").

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues