GE Healthcare · MRDiscovery2267863
Cannot Scan with Error 2267863 due to defective 9-pin cable 5165970 between XGA-J6 and current sensor (LEM).
Symptom
Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault.
Error message
2267863 The %axis XGA is reporting a I_Coil Wire Fault See Long Description Extended Error: The cables between the coil current sensor and the SB board is loose or has fallen off. Ensure cables J5 and J6 of the XGA are connected to the corresponding axis current sensor at the top of the Power Cabinet.
Applies to
- Models
- Discovery
- Product
- Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5304065 PGR Cabinet Signal Cable Kit - FRU 5304065 PGR Cabinet Signal Cable Kit - FRU
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.