Cannot Scan with Error 2267863 due to defective 9-pin cable 5165970 between XGA-J6 and current sensor (LEM).

Symptom

Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault.

Error message

2267863 The %axis XGA is reporting a I_Coil Wire Fault See Long Description Extended Error: The cables between the coil current sensor and the SB board is loose or has fallen off. Ensure cables J5 and J6 of the XGA are connected to the corresponding axis current sensor at the top of the Power Cabinet.

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5304065 PGR Cabinet Signal Cable Kit - FRU 5304065 PGR Cabinet Signal Cable Kit - FRU
Area
MR