Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26010009350164

LightSpeed RT16 & RT16 Xtra and Vision

Symptom

"Scan hardware Stopped Scan" with PDU errors "Scan hardware Stopped Scan" with PDU errorss

Error message

Error: 260100093 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time Mon Sep 5 13:55:52 2005 Error: 260100093 Host: tgp Process: tgp_subsys File: scIsrService.c Line: 246 Running diagnostic: kV and mA Test: Xray Functional Test Mon Sep 5 13:55:51 2005 Host : Table Gantry Processor Ermes # : 260100093 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 246 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Static Scan: 50164/1/1 PDU HVDC Buss Fault Status. The HVDC did not charge correctly in the alotted time. Threshold 1: > 200V in 200ms Threshold 2: > 500V in 1.2s. Possible causes: HVDC Cable Short Circuit, SlipRing, or Jedi Inverter.

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues