Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118011 CAN_WRITE_OVERRUN. This error occurs when the library waits more than CAN_DATA_MUTEX_TIMEOUT to be able to send a message See LD for complete error Fri Feb 3 07:59:35 2006 Error: 260118011 Host: orp Process: orp_subsys File: gen_exec.c Line: 459 Fri Feb 3 07:59:34 2006 Host : On-Board Rotating Processor Ermes # : 260118011 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 459 Function : SYSTEM COMMUNICATIONS : Generator CAN Serial Link Scan Type : None/Unknown Scan: 50075/6/0 CAN_WRITE_OVERRUN. This error occurs when the library waits more than CAN_DATA_MUTEX_TIMEOUT to be able to send a message.

Error: 260118315 JEDI ID Communication Timer Expired. Check power and commnication interconnects. See Long Description for complete error Mon Dec 6 12:04:49 2004 Error: 260118315 Host: orp Process: orp_subsys File: gen_exec.c Line: 1665 Mon Dec 6 12:05:01 2004 Host : On-Board Rotating Processor Ermes # : 260118315 Exception Class : Soft Severity : Pri/Soft File : gen_exec.c Line# : 1665 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 JEDI ID Communication Timer Expired. Check power and commnication interconnects.

Error: 2601

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.