Pular para o conteúdo principal
GE Healthcare · CTLightSpeed20010901220050200262920704

LightSpeed 4.X - Octane

Symptom

Unable to scan axial or helical; system cannot allocate image space as one chunk. Unable to scan axial or helical; system cannot allocate image space as one chunk

Error message

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request.

Error: 200502002 Cannot allocate image space as one chunk. See long description for error example Error: 200502002 Cannot allocate image space as one chunk Wed Mar 2 18:39:33 2005 Error: 200502002 Host: ct01 Process: scanRx File: srx_sma_manage_image_space.c Line: 682 Function: No system function reported Cannot allocate image space as one chunk. Total space of 62920704 (bytes), 2 paths returned.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.