GE Healthcare · MRExcite2248622223889722246662238905
RF amplifier not ready due to defective RF amp.
Symptom
Auto Prescan failed, RF Amplifier is not ready Getting another message that RF Amplifier is powering but message never goes away.
Error message
2248622 RF amplifier is not ready. See long description Error: 2248622 Host: HDx3T Process: NSP File: APS:aps_pass2 Line: 245 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: FSE-XL Coil: HDNV Array Scan: 4680/4 RF amplifier is not ready. 2238897 3T RF Amplifier High Voltage Power Supply Over Voltage (fault 06). The #2 PA High Voltage supply output voltage was excessive (7000DCV nom). Possible Causes: power supply defective. If fault persists, replace the amplifier. 2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready. 2238905 3T RF Amplifier PA High Voltage Supply Low (fault 15). Possible Causes: defective PA Tube, defective PA HV supply. If fault persists, replace the RF amplifier.
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 M5 Excite HDx HDi Platforms only.
- Component
- 2278810-4 3T 25KW RF AMPLIFIER KIT MKS ARX-X433 rev D 01480017ZZ RF POWER AMP model: 50-S26A-128 part#: 2278810-4
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.