GE Healthcare · MRDiscovery226742513951
2267425 No clock signal to IRF3 board error due to defective single strand fiber optic cable.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2267425 No Clock Signal to IRF3 board. Error: 2267425 Host: WIMRMR2 Process: diagCoord File: Irf3.cpp Line: 1402 Task ID: 0x493f6b0 Time: 13951 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A No Clock Signal to IRF3 board. The IRF3 board did not detect a clock signal on the J8 fiber optic clock cable. This condition will cause a TPS reset failure.-
Applies to
- Models
- Discovery
- Product
- Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.