GE Healthcare · CTLightSpeedVCT26013322426010102350444260007402260007453260001512
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
"Scanner Hardware Stopped Scan" with CAM or Filter errors
Error message
Error: 260133224 The collimator has detected motor slip during z-axis tracking. See long description Mon Nov 9 10:12:31 2009 Error: 260133224 Host: orp Process: orp subsys File: aperture z axis.cxx Line: 169 Mon Nov 9 10:12:30 2009 Host : Collimator Ermes # : 260133224 Exception Class : Secondary Severity : Pri/Soft File : aperture z axis.cxx Line# : 169 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 The collimator has detected motor slip during z-axis tracking. WARNING The cams will be re-referenced before the next scan.This may extend the ISD.Cam A Delta : -4323 encoder countsCam B Delta : -1 encoder counts Do Not Use - use on of the other two Error: 260101023 The (DCB or CCB) has notified the system of an error see long description for examples. Tue Dec 28 09:41:31 2004 Error: 260101023 Host: orp Process: orp_subsys File: rcib_erlog.c Line: 366 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260101023 Exception Class : Abort Severity : Primary File : rcib_erlog.c Line# : 366 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 The DCB has notified the system of an error condition which prevents scanning. See log for further inform
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5271284 VCT Collimator Motor and Coupling Assemb See long description for details Part #: 5271284 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: VCT Collimator Motor and Coupling Assembly FRU: Yes Status: Orderable Comments: Per ECR 2052055, 212008898, 4/7/08.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required