GE Healthcare · CTLightSpeed
LightSpeed 5.X (80/100kW)
Symptom
System fails to reconstruct images. System fails to reconstruct images
Error message
DARC /var/log/messages : timeout attempting to rsh to ig2 ( can be ig1 or ig3 ) See long description DARC /var/log/messages Aug 18 18:43:09 darc logger: is_IG_up 1 ig2 called Aug 18 18:43:09 darc is_IG_up:[recon_control:7973]: ig3 is up ( this will be typically 5 mins after start of DARC boot ) Aug 18 18:43:09 darc is_IG_up:[recon_control:7967]: ig1 is up Aug 18 18:43:10 darc is_IG_up:[recon_control:7970]: timeout attempting to rsh to ig2 ( this is ig2 failing to respond to DARC ) With no intervention DARC will retry to contact ig that failed to respond on fisrt attempt ( notice timestamps ) Aug 18 18:49:58 darc logger: is_IG_up 3 ig2 called Aug 18 18:49:58 darc mountd[3273]: authenticated mount request from ig2:855 for /usr/g (/usr/g) Aug 18 18:49:58 darc is_IG_up:[recon_control:7970]: ig2 is up ( ig2 up nearly 7 minutes after second start )
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2362872 Westville IG Module used on GOC3 and GOC4 consoles. See Long Description for details. 2362872 Westville IG Module used on GOC3 and GOC4 consoles. Part #: 2362872 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 19.84 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: IMAGE GENERATION COMPUTER - GRE FRU: Yes Status: Orderable Comments: **FE Assist** applies to CT Not Pet systems. If no stock is available order a Jarrell IG 5147443 to replace a Westville IG (2362872), you MUST order p/n 5193989 software release 06MW29.7, and perform a LFC. See Service note SN06101010A. PET must receive the Westville IG.(12/12/06 212001261) *** 2ND OPINION NOTES *** See the following URL for additional command line diagnostic troubleshooting: http://gemsvst4p.med.ge.com:7005/facets/CT/GRE%20command%20line%20diag%20test.pdf?
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required