Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18310250199183156

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Scanning aborted with Tube temp / pressure switch open - document in RFS - critical issue Scanning aborted with Tube temp / pressure switch open.

Error message

Error: 183102 Tube temperature/pressure switch open See the LD for the full error code Error: 183102 Tube temperature/pressure switch open Error: 183102 Host: OBC Process: Generator File: gen control.c 1.39 Line: 678 Function: XRAY GENERATION : Tube Pressure Sense Scan: 50199/1/4 Type: Axial Exception Level: Pri/Soft Time: 06:32:17:830 Log Series: 106 Tube temperature/pressure switch open.

Error: 183156 Unable to PREP generator. Sun Sep 5 11:10:28 2004 Error: 183156 Host: OBC Process: Generator File: gen_control.c 1.39 Line: 404 Function: XRAY GENERATION : Tube Rotor Control Scan: 4167/1/1 Type: Scout Exception Level: Pri/Most Severe Time: 11:10:29:170 Log Series: 43 Unable to PREP generator.

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-327160P1 12 Amp, 125Vac Dual Element Time delay fuse See long description for details Part #: 46-327160P1 Part Type: GP CONSUMABLES Part Weight: 0.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: 12.0A, 125VAC, DUAL ELEMENT TIME DLY FRU: Yes Status: Orderable Comments: FLS - USED ON AB SERVO AMP IN HSA. FUTS FW 26 '93. 12.0A, 125VAC, DUAL ELEMENT TIME DELAY.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues