GE Healthcare · MRDiscovery2268867560332256925
Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Defective Coil Sensor cable
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2268867 The gradient coil temperature cable is not detected at the SRI 2268867 Host: SCRB3MR Process: diagCoord File: SriManager Line: 3829 Diagnostic: SRI Board Level Diagnostics Iteration: 1 of 1 Action: 1 Task ID: 0xd0dcb58 Time: 56033 Object: SCP Function: Patient Handling: Control/Display The gradient coil temperature cable is not detected at the SRI. 2256925 StartupCondition error on Sri Can resulted in TPS reset failure
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- 5181092 Run M3344, MAG-SRI-J22 to MAG-BORE-G 5181092 Run M3344, MAG-SRI-J22 to MAG-BORE-G
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.