GE Healthcare · CTLightSpeedVCT23000090053852633554432123321812672703175000
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 230000900 WARNING slow write to disk See llong decsriptions for full error Thu Jan 29 08:54:34 2009 Error: 230000900 Host: darc Process: dataacq File: file_save_thread.cxx Line: 194 WARNING slow write to disk: 51.538526 sec to write 33554432 bytes; scan CT07/152/3/3-1233218126/727031; scan seq 75000 Example: This took 51 seconds to write 33 mb. which is extremely long. 51.538526 sec to write 33554432 bytes; Error: 245274 WARNING- Scan setup took 3159 ms overall. See Long Description for more details Thu Mar 23 07:00:44 2006 Error: 245274 Host: fhtc2 Process: scanmgr File: scanmgr lib.cxx Line: 1322 Function: System Control : Scan Control Scan Type: Cine Scan: 0/0/0 Scan Seq Id: 17000 WARNING- Scan setup took 3159 ms overall. This could potentially cause scan timing issue or scan abort. Fri May 9 10:50:31 2008 Error: 245272 Host: ct01 Process: scanmgr File: scanmgr_lib.cxx Line: 1189 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 2779000 WARNING- Scan setup took 1897 ms on Step:7 (Get Cal Module).
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Software
- 09MW08.10 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) June 2009 Inital M3 release. This software is now obsolete, this site should be upgraded to 11HW12.5 software released via FMI 25427 09MW08.10 for GOC6 LightSpeed VCT 7.2 (64 Slice Only) June 2009 Inital M3 release. LightSpeed 7.2 ( 09MW08.10 ) Software Release supports GOC6 consoles used on VCT 64, VCT XT, VCT XTe and VCT Std. - LightSpeed 7.2 (09MW08.10 )Software for GOC6 Collector 5328048 Rev 3 Includes OS and Application CD's - Operating System Linux 5.3.8 Software CD 5323584 Rev 1 - Application LightSpeed VCT GOC6 DVD 5336533 Rev 3 - Class C LightSpeed 7.2 5336535 Rev 2 - Class M LightSpeed 7.2 5336536 Rev 2 An On-site/Remote identification aid . (Example: check_config) Result from check_config or swhwinfo 09MW08.10.V40_H_H64_G_GTM 09MW08.10.V40_H_H64_G_GTM Please create an ITRAK report for any software related issues
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.