Will Not Scan With Gradient Errors due to PS-Lite2 Control Board.

Symptom

Will Not Scan with Gradient Errors.

Error message

2224666 Auto Prescan failed, Master Gradient is not ready

2244860 The GP is reporting a SGA Power Supply OverCurrent Fault. Same as 2244860 and 2244590. 2244860 errorcode on Post 14M5 is the same error as 2244590 on 11X, G3 and 2244591 on 12X, E2, 14M4 (HDx, HDe).2244860 "The GP is reporting a SGA Power Supply OverCurrent Fault.

2244858 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage. the same error as 2244593 on 11X, G3 and 2244594 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244858 error code on Post 14M5 is the same error as 2244593 on 11X, G3 and 2244594 on 12X, E2, 14M4 (HDx, HDe), and openMR4,5. 2244858 The GP is reporting SGA Power Supply Internal Wiring Fault or Control Undervoltage.

2244859 The GP is reporting a SGA Power Supply OverTemperature Fault. System will wait five minutes before attempting recovery. Same as 2244590. 2244859 errorcode on Post 14M5 is the same error as 2244590 on 11X, G3 and 2244590 on 12X, E2, 14M4 (HDx, HDe). 2244859 The GP is reporting a SGA Power Supply OverTemperature Fault. System will wait five minutes before attempting recovery. 2244859 errorcode on Post 14M5 is the same error as 2244590 on 11X, G3 and 2

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 15.0 For Hino Excite HDe 15.0 systems only.
Component
2336029 P9438HB CONTROL BOARD ASSY
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.