GE Healthcare · MRSigna224828022494182249307108193483020000
Unable to download or reset TPS With DRF Error Due to DRF.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2248280 DRF board is not available Please make sure DRF is plugged in the MGD chassis. 2249418 Network problem detected. Check MGD subnet network configuration. See Long Description Wed Apr 14 09:27:10 2004 2249418 2 CARD diagCoord mgdArp.c 157 Task ID: 0xf99df8 Time: 981 Object: SCP Network problem detected. Check MGD subnet network configuration. etherAddrResolve failed for network interface motfec0, IP 10.47.1.2, numTries 5, delay 100 ticks. status = -1. 2249307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Error seen with MGD Diags: 1081934830 0 1 Wed Apr 14 09:27:10 2004 2249307 2 CARD diagCoord pearlMap.c 1806 Task ID: 0x0 Time: 0 Object: SCP Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Value 0xAAAA not found at address 0xC03010A8 after 20000 msec.->
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2298332 DIGITAL RECEIVER AND FILTER II DRF2 2298332 DIGITAL RECEIVER AND FILTER II DRF2 Part #: 2298332Part Type: GP CONSUMABLES Part Weight: -3175168.25 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: DIGITAL RECEIVER AND FILTER II DRF2FRU: YesStatus: OrderableICV/Transfer Price: 858.16 USD /Each ICV/Transfer Price: 858.16 USD /Each List Price: 8939.61 USD /EACH
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.