Multi Coil Bias Short Circuit Faults Due to loose pin in coil connector.

Symptom

Unable to Scan With Multi Coil Bias Faults Unable to scan with Multi Coil Bias Faults

Error message

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2252303 Multicoil Bias Fault Detected, Short Circuit. Channel X of Y See Long Description Error: 2252303 Host: lx-mr Process: NSP File: SCP:DCB Line: 1030 Multicoil Bias Fault Detected - Short Circuit. Try re-seating all coil connectors. MCD1 Channel 3 of 16, counting from 1 Value: -310 mA Limit: -301 mA Mask of failed channels: 0x3c Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. A short circuit fault occurs when the actual value is more negative than the pre-defined limit. This could be caused by a shorted diode in the coil or a short elsewhere in the system. Check coil, system cables, connectors and configurations. The mask of all failed channels indicates additional channels with the same fault - convert the hex number to a bitmap to decipher the mask, with channel 1 indicated by the least significant bit and channel 16 indicated by the most significant bit. For exa

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR