Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT210000428780138242688168521000040721000044735841495

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Unable to display any images; exam browser listing in image works is empty and fails to refresh.

Error message

Error: 210000428 Unable to establish connection for repository iosterra|terradb session See long description for full error Tue Jan 29 06:51:29 2008 Error: 210000428 Host: gevct2 Process: dms File: Initialize.cxx Line: 396 Unable to establish connection for repository iosterra|terradb session

Buffer I/O error on device md0, logical block 78013824 seen on HP or DARC in /var/log/messages See LD for complete error Buffer I/O error on device md0, logical block 78013824 cd /var/log tail –1000 messages | more Look for any disk drive errors like these below in the HP or the DARC messages log: 16:13:48 ctcwct01 kernel: Buffer I/O error on device md0, logical block 26881685 Aug 2 12:11:33 darc kernel: Buffer I/O error on device md0, logical block 78013824

Error: 210000407 Image space query failed. See Long Description for more details Mon Feb 27 07:19:00 2006 Error: 210000407 Host: csbipct2 Process: dms File: Initialize.cxx Line: 1085 Image space query failed. Error code: -1

Error: 210000447 unable to connect to session dms_session with error code -2 See long description for details Tue Jan 29 06:51:29 2008 Error: 210000447 Host: gevct2 Process: dms File: DMFInstall.cxx Line: 229 Error: 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5117866-5 73 GB HDD used in HPxw8200 GRE. See Long Discription for Details. Part #: 5117866-5 Part Type: GP CONSUMABLES Part Weight: -139998.60 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73 GB HDD for Lighting System Controller 5117866-2 FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues