Pular para o conteúdo principal
GE Healthcare · MRSigna2224666

Will Not Scan With Patient Comfort Sensor Trip Due to Scan room temperature was warmer than normal.Customer Facilities E…

Will Not Scan With Patient Comfort Sensor Trip Due to Scan room temperature was warmer than normal.Customer Facilities Engineer must correct the Scan Room temperature issues wqr

Symptom

Will Not Scan With Patient Comfort Sensor Trip. See Long Description Patient comfort sensor trip; cannot scan due to bore sensor indicating temperature is too hot for patient safety. A message "Patient Comfort Sensor Trip" appears in the message box on the left side of the scan page at the start of autoprescan.

Error message

2224666 Auto Prescan failed, Patient comfort temperature sensor trip. Mon Apr 5 07:48:24 2004 Error: 2224666 Host: lxmr3 Process: psc File: PscMainControl.cc Line: 1334 Auto Prescan failed, Patient comfort temperature sensor trip

Applies to

Models
Signa
Product
Signa HDxT Apps. For use with HD16.0_VO1 and HD16.0_VO2 version software only. For use with HD16.0_VO1 and HD16.0_VO2 version software only.
Software
HD16 Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only. Includes HD16.0_VO1 and HD16.0_VO2 version software and applies to 1.5T and 3.0T "Signa HDxT Apps" systems only.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues